22
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202522A
Replacing the Fuses
Service frequency
As needed.
Caution
Use proper Electrostatic Discharge (ESD) procedures when performing this maintenance. Failure to use
proper ESD preventative measures could cause damage to the control board.
Tools/equipment
• Small Phillips screwdriver
• Medium Phillips screwdriver
• Needle-nose pliers
• ESD strap
• Model 750 Control Board
• Flat screwdriver (optional)
Method
1. Detach the hose (see Detaching Hose on page20 for instructions).
Disassembling the Temperature Management Unit
1. Using a medium Phillips screwdriver, remove the two screws that attach the IV pole clamp or plug.
2. Turn the unit so the bottom of the unit faces up. Remove the four bottom screws with a medium
Phillips screwdriver.
3. Unplug the 2-pin black connector used for the main voltage.
4. Using the medium Phillips screwdriver, remove the two screws that connect the 2 ground wires.
Remove the bottom enclosure.