Bair Hugger Model 750 Temperature Management Unit Service Manual
English 202522A
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202522A
Fault Conditions
e Model 750 temperature management unit’s soware recognizes several nonhazardous fault conditions
including:
• Sensor failures
• Heater failures
• Computer failure
• Soware failure
• Keypad failure
• Power supply failure
• Blower motor failure
• Forced over-temperature test failure
When a fault condition exists, the following events occur:
• Fault indicator light ashes.
• Audible alarm sounds.
• Blower heater turns o.
• Fault code appears in the alphanumeric display.
• Up to 5 fault codes are stored in memory.
When the audible alarm sounds, press any button on the keypad to temporarily silence it. e temperature
management unit must be unplugged before it can be restarted. See Viewing the Fault Code Log and the
Fault Code Table for more information about specic fault codes.
Over-Temperature Condition
When an overtemperature condition exists, the following events occur:
• Over-temp indicator light ashes.
• Audible alarm sounds.
• Blower and heater turn o.
• Fault code appears in the alphanumeric display (see Fault Code Table).
When the audible alarm sounds, press any button on the keypad to temporarily silence it. e temperature
management unit must be unplugged before it can be restarted.
* Up to 3 fault codes in units with < Rev L soware. e Rev version appears in the display window upon
connection to a power source.