English 202430A
17
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202430A
Normal Operating Temperature Testing
WARNING: Perform all testing and adjustments of warming
unit temperature a Temperature Test Unit Model 22110.
Arizant Healthcare Inc. assumes no responsibility for the reliability,
safety, or performance of the Bair Hugger system if warming unit
temperature testing or adjustments are made in any other manner
than that described here. Improper measurement or adjustment of
the warming unit’s normal operating temperature could result in
patient exposure to temperatures outside of the indicated range,
and may lead to patient injury.
Service Frequency
6-month intervals, or 500 hours of use
Tools/Equipment
Temperature Test Unit Model 22110
Method
1. Insert the open hose end from the warming unit in the port
on the temperature test device (see Figure 6).
NOTE: Do not block the rear air vents on the test device.
2. Turn the warming unit power ON and select the
HIGH temperature setting. Always begin testing at the
HIGH setting.
3. Note the ambient room temperature.
4. After approximately 5 minutes of operation, note the
temperature read-out. Use Figure 8 to correlate ambient
temperature to the warming unit’s operating temperature.
At HIGH, Nominal Temperature on Test
Kit (°F)
Ambient Room Temperature (°C)
At HIGH, Nominal Temperature on
Test Kit (°C)
Figure 8. Room Temperature vs. High
Temperature Output
Warming unit temperature setting should be on HIGH when
conducting this test
1
Ambient Room Temperature (°F)