Bair Hugger Model 505 Service Manual
18
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202430A
Normal Operating Temperature Testing
5. After correcting for ambient temperature, the tested normal
operating temperature should meet the limits specified in
Table 2.
6. If the HIGH temperature setting is not within specified
limits, adjust the normal operating temperature. See the
Normal Operating Temperature Adjustment section.
7. If the HIGH temperature setting is within specified
limits, verify the MED and then LOW settings against
specified limits (allow the warming unit to operate at each
temperature setting for about 3 to 5 minutes
before test).
8. Turn the warming unit to STANDBY. Remove the hose
end from the port on the temperature test device.
9. Record the maintenance action taken and the hours of
operation (read from the hour meter) in the Maintenance
Log section.
Table 2. Operating Temperatures Model 505
(Operating temperature = Average temperature at the end of the hose.)
HIGH: 43° ± 3°C (109.4° ± 5.4°F)
MED: 38° ± 3°C (100.4° ± 5.4°F)
LOW: 32° ± 3°C (89.6° ± 5.4°F)
Note: Air temperature reaching the patient is approximately 2°C lower than
the listed temperatures.