Troubleshooting
e model 775 soware routinely checks unusual operating conditions� If such a condition occurs, the
system may trigger a fault condition and alarm� e warming unit can be reset by disconnecting the
warming unit from the main power source, allowing 30-60 seconds for the system to reset� If the fault
condition reoccurs upon connection to the main power source, the unit may have a hard-fault or permanent
error�
Call Technical Support for Replacement Parts at 1-800-733-7775�
Fault Code Possible Cause Solution
Permanent errors – a fault condition always occurs at turn on or after pressing a temperature setting
FC-001
FC-002
FC-003
FC-004
FC-005
FC-006
FC-007
Faulty hose end sensor at
patient end�
Replace hose� Refer to service manual or IFU
included with replacement part for instructions�
Faulty elbow sensor� Replace elbow sensor� Refer to IFU included with
replacement part for instructions�
Hose disconnected from
unit�
Reattach hose�
Calibrate and return to service�
FC-008
FC-009
FC-010
Faulty heater element or
faulty control board�
•Verifyfanrunsinambient.
•CheckcontinuityofJ4connectoroncontrol
board�
White to black
White to blue
White to yellow
If “open”, replace heater� Otherwise, replace
control board�
Calibrate and return to service�
FC-050 Defective user interface� Replace user interface�
Calibrate and return to service�
FC-053, 054 EEPROM error�
Reset microprocessor�
(Press the Bair Hugger logo and Standby button
at same time while connecting to power source�
Release buttons when Standby light remains on)�
Calibrate and return to service�
FC-100
FC-101
Defective control board� Replace control board�
Calibrate and return to service�