Bair Hugger Model 775 Temperature Management Unit Service Manual
English 202526A
Check the 3M
TM
Bair Hugger
TM
therapy website to ensure you have the most recent version of this document. www.bairhugger.com reorder #202526A
CAUTIONS
1� Except for specic blanket models, Bair Hugger blankets are not sterile and are all intended for single
patient use ONLY� Placing a sheet between the Bair Hugger blanket and the patient does not prevent
contamination of this product
2� Monitor the temperature and cutaneous response of patients who are incapable of reacting,
communicating and/or who are without a sense of feeling every 10-20 minutes or according
to institutional protocol� Monitor the patient’s vital signs regularly� Adjust air temperature or
discontinue therapy when the therapeutic goal is reached or if vital sign instability occurs� Notify
physician of vital sign instability immediately
3� Do not leave pediatric patients unattended during therapy
4� Do not initiate temperature management therapy unless the temperature management unit is free
from mechanical damage and is safely placed on a hard surface or securely mounted� Otherwise,
injury may result
5� To prevent tipping, clamp the Model 775 temperature management unit to an IV pole at a height that
provides stability� We recommend clamping the unit no higher than 44" (112 cm) on an IV pole with a
minimum 28" (71 cm) diameter wheelbase� Failure to do so may result in IV pole tipping, catheter site
trauma, and patient injury
6� Electrical shock hazard� Do not disassemble the temperature management unit unless you are a
qualied service technician� ere are electrically live parts within the unit when it is connected to a
power source, even when the unit is in Standby mode
7� To reduce the risks associated with environmental contamination follow applicable regulations when
disposing of this device or any of its electronic components
NOTICES
1� e Bair Hugger temperature management unit meets medical electronic interference requirements�
If radio frequency interference with other equipment should occur, connect the unit to a dierent
power source�
2� Federal law (USA) restricts this device to sale by or on the order of a licensed healthcare professional
3� To avoid Bair Hugger warming unit damage:
• do not immerse the Bair Hugger warming unit of warming unit parts of accessories in any liquid
or subject them to any sterilization process
• do not use solvents such as acetone or thinner to clean the warming unit; avoid abrasive cleaners
• clean warming unit exterior with so cloth using plain water or a mild, all-purpose or
nonabrasive cleaner�