Replacing the Air Filter
Service frequency
Every 12 months or 500 hours of use�
Tools/equipment
• Replacement air lter
• #2 Phillips screwdriver
WARNING
Do not attempt to clean the air lter as it may contain hazardous microorganisms� Discard the lter in a
manner consistent with institutional protocol for biohazardous material
Method
1� Disconnect the temperature management unit from the grounded power source�
2� Turn the temperature management unit upside down�
3� Remove the 2 screws located in the louvered lter cover
4� Li the lter cover o the temperature management unit
5� Remove the air lter and discard it�
6� Place the new lter in the lter compartment with the gasket toward the lter ledge�
7� Replace the lter cover with the louvers facing the back of the temperature management unit�
8� Replace the 2 screws in the lter cover